OPERATING INSTRUCTIONS
This section contains detailed procedures for the operation of the Dynatron 900+ Traction Unit. These
procedures are designed to familiarize the user with the operation and function of the device. Read this
section thoroughly along with all contraindications, warnings, and precautions before operating this device.
WARNING
• Thoracic, Pelvic and Cervical harnesses must be securely in place before traction treatment
begins.
• Always STOP a treatment before removing, or adjusting harnesses.
• Check the traction cable for knots, twisting, fraying and damage of any kind.
• Never use worn or damaged harnesses as these may result in injury to the patient or lessen
the effectiveness of a treatment.
• See the contraindications, warnings, and precautions in this manual before administering a
treatment to a patient.
• The patient controlled REMOTE EMERGENCY STOP cable must be inserted into the front of
the Dynatron 900+
device before powering-on the device
.
Prior to treatment, prepare all belts and harnesses according to manufactures directions and carefully
position the patient.
CERVICAL
,
THORACIC AND PELVIC HARNESSES
.
The cervical, thoracic and pelvic harnesses
must be properly fitted according to manufacturers’ instructions. Belts should be securely
tightened without creating discomfort. Harnesses secure the patient to the traction device while
supporting the musculature, thus allowing for the proper application of the distraction forces.
€
POSITIONING THE PATIENT AND TABLE
.
As the effect and effectiveness of traction are related
to the position of the body as well as the force and duration of pull, make sure that the
positioning of the patient and the table are according to prescribed protocols. The use of
pillows or bolsters under the arms, head, knees, or ankles may be necessary to achieve proper
positioning. See published literature.
QUICK SETUP INSTRUCTIONS
1.
Prepare the patient for the traction treatment. Make sure all belts and harnesses are securely attached.
2.
Plug in the remote stop cable and hand it to the patient.
3.
Turn on the Dynatron 900+ device.
4.
Set Treatment Time:
(1-99 min.)
5.
Set Hold Force Timer:
(1-99 sec.)
6.
Set Rest Force Timer:
(1-99 sec.)
7.
Set Hold Force:
(4 – 200 lbs. or 2 - 91 kg.)
8.
Set Rest Force:
(0 – 197 lbs. or 0 - 90 kg.) Must be set lower than Hold Force value.
9.
Adjust Speed:
Range: Slow to Fast
10.
Press START