Dynatron 900
+™
4
DESCRIPTION OF PHYSICAL FEATURES
1.
ON
/
OFF TOGGLE SWITCH
The On/Off Toggle Switch is located on the left-side of the lower front panel. When the toggle switch
is in the “1” (on) position, the LED above the Reset Key is illuminated green as well as other display
windows and indicators on the device faceplate.
2.
TIME SELECTOR ARROWS
Located in the upper right-hand corner of the faceplate, these Time Selector Arrow Keys are used to set
the treatment time within a range of 1-99 minutes. As the keys are pressed, the time appears in the
Time Display Window and will count down during treatment.
3.
START KEY
When the Start Key is pressed, treatment begins. The time automatically begins to count down and the
green LED above the Start Key is illuminated. When the treatment is complete, a tone will sound
signaling that the treatment is finished and the green LED will go off.
4.
RESET KEY
The Reset Key has four functions: 1. Press the Reset Key before beginning a new treatment to clear all
previous settings. 2. If treatment is interrupted, immediately push the Reset Key to release the tension
on the rope and end the treatment. 3. Used in combination with the SP Key to select Max Force.
4. Concludes treatment and releases all tension from the traction rope.
5.
TIME DISPLAY WINDOW
Located in the upper right corner of the device faceplate, the digital readout in the Time Display
Window counts down the preset treatment time. Treatment time is set by using the Time Selector
Arrow Keys below the Time Display Window.
6.
REMOTE STOP JACK
The Remote Stop is controlled by the patient and can be used to terminate the treatment at any time.
The Remote Stop should be plugged into the device prior to turning on the device.
7.
SP KEY
Located toward the left side of the control panel, the SP Key toggles through four settings that control
the volume of the audible signals: Mute/Low/Mid/Max.
8.
“
N
/
KG
/
LB
”
KEY
This key is located toward the right side of the control panel. It allows the user to control the unit of
measurement in which the Traction Force will be displayed: N=Newtons; kg.= kilograms; lb.= pounds.
As the key is pressed, the green LED next to the selection will be illuminated indicating that the
selection has been activated.
9
.
SP PLUS RESET KEY
While holding down the Reset Key, use the SP Key to select the Max Force range. The green LED
next to each selection will be illumined when the selection is activated.
890N / 91kg / 200lb
490N / 50kg / 110lb
180N / 18kg / 40lb
10.
TRACTION FORCE DISPLAY WINDOW
(
POUNDS
/
KILOGRAMS
/
NEWTONS
)
The LEDs in the Traction Force Display Window indicate the pounds, kilograms or Newtons of traction
force during both "Release" and "Hold" cycles. The display counts up as the traction system releases or
pulls. The target number will remain illuminated during the “Hold” or “Rest” time cycles. The active
cycle (Hold or Rest) is indicated by the illuminated arrow pointing toward the “Hold” or “Rest” display
windows located in the center of the control panel.
When selecting Hold or Rest forces using the force selection knobs, the Traction Force Display Window
will briefly show the Hold or Rest force dialed. The displayed selection will hold for two seconds before
the display automatically returns to “000.” In “Reset” mode, the digital display reads “'000.” Traction
forces may be changed during treatment; however the Start Key must be pressed after each change.