18.
MAX FORCE LEDS
Under the Max Force heading, three maximum force selections are listed. Selection is made by holding
down the “Rest” Key while pressing the SP Key. The LED next to each Max Force selection becomes
illuminated when the selection is active.
890N / 91kg / 200lb
490N / 50kg / 110lb
180N / 18kg / 40lb
19.
SPEED CONTROL KNOB
This knob labeled “Speed” controls the variable "Pull" and "Release" speed of the rope. The rope
speed is increased by turning the knob to the right and reduced by turning the knob to the left. Rope
Speed can be adjusted on a gradient scale of slow to fast (8 lbs.-17 lbs. per second).
20.
POWER SOCKET
Plug the Dynatron 900+ power cord firmly into the power socket located on the back of the device
prior to plugging the device into a wall socket. Do not place the cord or the device in a place where the
cord could be tripped over or accidentally pulled out of its socket during a treatment.
21.
TRACTION PULLEY
The Traction Pulley can be rotated to any desired position.
22.
TRACTION CORD
The traction rope is connected to the spring loaded clip by a knot.
23.
SPRING LOADED TRACTION CLIP
The traction Clip is used to attach the traction belts to the traction
device. When operating the device, the traction clip should be
situated with the clip opening facing up as illustrated in the
diagram.
24.
FUSES
The Dynatron 900+ requires two fuses:
F 1.6A 250V (100V, 110V, 120V)
F 0.8A 250V (220V, 230V, 240V)
SAFETY SYSTEMS
Warning Alarms will sound under the following conditions:
1.
Overload (Hold Force setting is greater than maximum value of the selected range).
2.
Traction force exceeds preset force by approximately 8 lbs.
3.
Remote Switch is pressed
4.
Rest Force setting is equal or greater than the Pull Force setting.
5.
Device malfunctions
6.
Traction force exceeds 200±4lb. (91±2kg /890±20N)