Dynatron 900

+™

 

 

DETAILED TREATMENT SETUP

 

1.

 

According to manufactures’ instructions, prepare the patient for treatment.  Make sure all belts 
and harnesses are securely attached. 

 

2.   The 

Remote Stop

 cable must be plugged into the console before the device is turned “on.”  

Connect the Remote Stop plug to the Remote Jack located in the center of the rear panel and 
hand the Remote Stop switch to the patient.  

 

3.  

Turn the Power Switch to “1” (on).

  The On/Off Toggle switch is located on the lower 

left-side of the control panel.  When the toggle switch is in the “1” (on) position, the LED above 
the Reset Key is illuminated green.  The digital display windows are also illuminated green 
indicating that there is power to the device.    

 

4.

 

Set Treatment Time: (1-99 min.)  

Using the

 

up/down arrow keys located immediately under 

the Time Display Window, set the treatment time.  The treatment time will appear in the Time 
Display Window and the display will count down during a treatment.  Treatment time ranges 
from 1-99 minutes and may be adjusted in 1 minute intervals.  

 

Note:  Traction forces may be changed during treatment; however the Start Key must be pressed 
 

  after each change.  Changes made during the treatment, will not take effect until the next new 
  Hold/Rest cycle begins.   

 

4.

 

Set Hold Force Time (1-99 sec.).

  Located directly above the “Hold Force” knob, Hold Time 

is selected by pressing the up/down arrow keys below the Hold Display.  This selection 
determines the amount of time the traction tension will hold after reaching the set Hold Force.  
Hold Force Time ranges from 1 - 99 seconds and may be adjusted in 1 second intervals.  When 
a Static traction treatment is desired, set the Hold Time at “- -”.   

 

5.

 

 

Set Rest Force Time (1-99 sec.).  

Located directly above the “Rest Force” knob, Rest Time is 

selected by pressing the up/down arrow keys below the Rest Display.  This timer determines the 
amount of time the Rest tension will hold after reaching the preset Rest Force.  Rest Force Time 
ranges from 0 - 99 seconds and may be adjusted in 1 second intervals.  When a static traction 
treatment is desired, set the Hold Time at “- -”.  The Rest Force setting is ignored. 

 

6.

 

Set Hold Force (

4 – 200 lbs.; 2 – 91 kg; or 20 – 890N).

  

Hold Force is set by using the 

circular knob located on the right-hand side of the control panel.  Three numbered circles 
surround the knob, the inside circle indicates force in Netwons, the middle circle in kilograms, 
while the outside circle indicates force in pounds.  Force is selected by aligning the gray dot on 
top of the knob with the desired/ pounds/kilograms of force indicated on the circles.  Hold force 
ranges from 

4 – 200 lbs.; 2 – 91 kg; or 20 – 890N.  

 

 

7.

 

Set Rest Force (

0 – 197 lb.; 0 – 90 kg; 0 – 890N).

  

Rest force is set by using the circular knob 

located on the left-hand side of the control panel.  Three numbered circles surround the knob, 
the inside circle indicates force in Netwons, the middle circle in kilograms, while the outside 
circle indicates force in pounds.  Rest Force is selected by aligning the gray dot on top of the 
knob with the desired/ pounds/kilograms of force indicated on the circles.  Rest force ranges 
from 

0 – 197 lb.; 0 – 90 kg; 0 – 890N.  

 

 

9.

  Adjust Speed

 

(

8 lbs.-17 lbs. per second).  Adjust the variable "Pull" and "Release" speed of the 

rope by using the Speed knob located on the left side of the control panel.  The rope speed is 
increased by turning the knob to the right and reduced by turning the knob to the left.  Rope 
Speed can be adjusted between slow and fast (8 lbs. -17 lbs. per second).   

    

10. 

Press Start.

  Press Start and the treatment begins.  If a beeper sounds when Start is pressed at 

the beginning of a treatment, be sure that the “Rest” Force is set lower than the “Hold” Force.  
Press the Reset Key to reset the alarm and indicator lamps.  

Содержание 900+

Страница 1: ......

Страница 2: ...imes DANGER Explosion Hazard Do not use in the presence of flammable gas or anesthetics Do not use the Traction System near X ray or Diathermy devices These devices may emit high frequency noise that could affect the operation of the Dynatron 900 Traction device INDICATIONS FOR USE Dynatron 900 Traction System The Dynatron 900 Traction system is intended for medical purposes for use in conjunction...

Страница 3: ... SPINAL DECOMPRESSION 13 CONTRAINDICATIONS 13 PRECAUTIONS WARNINGS 13 SECTION V TECHNICAL INFORMATION 14 SUGGESTED MAINTENANCE SCHEDULE 14 SERVICE TO BE PERFORMED BY A TECHNICIAN 14 MAINTENANCE PERFORMED BY USER 14 CAUTION 14 WARNING 14 RETURNING A UNIT FOR REPAIR 15 PACKAGING AND SHIPPING 15 CARE AND CLEANING INSTRUCTIONS 15 ENVIRONMENTAL CONDITIONS 15 SECTION VI ASSEMBLY AND MAINTENANCE 16 DYNAT...

Страница 4: ...Dynatron 900 iv ...

Страница 5: ...ructions for the Dynatron 900 Microprocessor controlled Traction system A copy of this manual should be left with the device at all times BEFORE YOU TREAT A PATIENT Before administering a treatment to a patient with the Dynatron 900 familiarize yourself with all the operating instructions as well as the contraindications warnings and precautions for this device After reading the information provid...

Страница 6: ...re covered by the warranty policy Warranty registration must include the device serial numbers Read the operating instructions in this manual carefully before proceeding with a treatment Connect the AC power cord which is equipped with a hospital grade UL listed plug to a properly grounded 110 120V 60 Hz AC 220V 60 Hz if applicable outlet The power cord must be firmly plugged into the Dynatron 900...

Страница 7: ...0N 200 lbs approximately 91kg 890N Note 5 lbs 2 5kg 25N for all settings Traction Force Compensation Computerized Automatic Compensation Force Units Pounds Kilograms or Newtons Hold Rest Time 1 99 sec independently set Treatment Time 1 99 min 2 Traction Speed 9 35 sec 0 200 lbs variable speed motor stepless continuous and adjustable Continuous Traction Release Time 00 99 seconds by static traction...

Страница 8: ... at any time The Remote Stop should be plugged into the device prior to turning on the device 7 SP KEY Located toward the left side of the control panel the SP Key toggles through four settings that control the volume of the audible signals Mute Low Mid Max 8 N KG LB KEY This key is located toward the right side of the control panel It allows the user to control the unit of measurement in which th...

Страница 9: ...Dynatron 900 5 DYNATRON 900 NUMBERED FEATURES ...

Страница 10: ... above the Rest Force knob Rest Time is selected by pressing the arrow selector keys under the Rest Display Window This display indicates the amount of time the rope tension will Rest following the release phase of the traction cycle Rest force time ranges from 1 99 seconds and may be adjusted in 1 second intervals When a static traction treatment is desired press the down the Rest Force Arrow Key...

Страница 11: ...the device in a place where the cord could be tripped over or accidentally pulled out of its socket during a treatment 21 TRACTION PULLEY The Traction Pulley can be rotated to any desired position 22 TRACTION CORD The traction rope is connected to the spring loaded clip by a knot 23 SPRING LOADED TRACTION CLIP The traction Clip is used to attach the traction belts to the traction device When opera...

Страница 12: ...d carefully position the patient CERVICAL THORACIC AND PELVIC HARNESSES The cervical thoracic and pelvic harnesses must be properly fitted according to manufacturers instructions Belts should be securely tightened without creating discomfort Harnesses secure the patient to the traction device while supporting the musculature thus allowing for the proper application of the distraction forces POSITI...

Страница 13: ...me is selected by pressing the up down arrow keys below the Rest Display This timer determines the amount of time the Rest tension will hold after reaching the preset Rest Force Rest Force Time ranges from 0 99 seconds and may be adjusted in 1 second intervals When a static traction treatment is desired set the Hold Time at The Rest Force setting is ignored 6 Set Hold Force 4 200 lbs 2 91 kg or 20...

Страница 14: ...LED above the Rest key will flash the Treatment Time Display Window will flash 00 and the tension on the rope will ramp down to 4 lbs Press the Reset Key to conclude the treatment and release all tension from the rope The 900 is now ready for program changes or to setup a new treatment ...

Страница 15: ... hand it to the patient 3 Turn on the Dynatron 900 device 4 Set Treatment Time 1 99 min 5 Set Hold Force Timer 1 99 sec 6 Set Rest Force Timer 1 99 sec 7 Set Hold Force 4 200 lbs or 2 91 kg 8 Set Rest Force 0 197 lbs or 0 90 kg Must be set lower than Hold Force value 9 Adjust Speed Fast Intermittent Cycle Slower Intermittent Cycle with Longer Ramp Time 10 Press START HARMONIC INSTRUCTIONS 1 Prepar...

Страница 16: ... be set lower than Hold Force value 14 Adjust Speed Slow minimum setting 15 Press START NOTE AS A SAFETY PRECAUTION IT IS NECESSARY TO RESET BOTH THE HOLD AND REST FORCES WHENEVER THE POWER SWITCH IS ARE TURNED ON FAILURE TO DO SO WILL RENDER THE DEVICE INOPERABLE SPECIAL USAGE PROCEDURES ADJUST HOLD REST FORCES AFTER ATREATMENT IS IN PROGRESS 1 Once the initial Hold Force is reached during a trea...

Страница 17: ...joint and or soft tissue injury Any condition that grows worse after treatment The presence of vascular compromise PRECAUTIONS WARNINGS Ligamentous Strains and Joint Hypermobility Patient s with Traction Anxiety causing muscles to become over tense and or cardiac respiratory insufficiencies Discontinue treatment immediately if there is an increase in symptoms or pain or if paresthesia prickling or...

Страница 18: ... to the unit and may endanger the life of the patient The unit may only be opened and serviced by authorized personnel Unplug the unit before opening Dynatronics recommends that the Dynatron 900 be calibrated annually to ensure the unit is working at peak performance When disposing of a Dynatron 900 unit comply with the laws and procedures required in your area according to directives in ISO 14000...

Страница 19: ...or device CARE AND CLEANING INSTRUCTIONS Clean the outer surface of the Dynatron 900 device with a slightly damp or lightly moistened cloth Mild household cleaners work well on the frame Do not spray the solution directly on the unit First moisten the cloth and then wipe the unit off Solvents caustic solutions or harsh or abrasive cleaners must never be used Avoid stretching electrical cords to fu...

Страница 20: ...ced or removed from the device turn the pulley head to the left until the breach points are aligned and the pulley easily slips from the enclosure Do not force REMOVING THE CHASSIS 1 Remove the two screws from the back panel of the chassis A 2 Remove the bottom plate and side panels will fall free ...

Страница 21: ...e device with hand screws E DYNATRON 900 PLATFORM INSTALLATION 1 Align hole A on the machine with pin F on the installation plate 2 Turn the machine to the desired orientation 3 Tighten the 4 hand screws E DYNATRON 900 TABLE INSTALLATION 1 Tighten stud to two screw holes G 2 Attach universal mounting bracket to stud 3 Tighten the knobs Orientation Options ...

Страница 22: ...HINED HOLE SIZES A Alignment hole 3 8 dia B Alignment set pin holes dia C Screw holes for table mount wall mount 5 16 18UNC D Knob for wall bracket 5 16 18UNC E Screw holes for stand mount or mounting bracket 5 16 18UNC Device Front ...

Страница 23: ...ble through the hole of pulley B go around the main shaft and under cable pull cable tight to secure the end of cable against the main shaft Fig 1 4 Pass side P2 of cable to pulley C Fig 2 5 Then pass pulley D Fig 3 6 Go around head pulley Fig 4 7 Tie the hook with cable as shown in Fig 5 7 8 En circle the cable on pulley B 9 Turn pulley B until the keyway faces you Fig 8 10 Turning the rotary spr...

Страница 24: ...sert the new fuse STEP 4 Insert the fuse case into the fuse holder STEP 5 Turning clockwise use a screwdriver to reattach the fuse cover FUSE 250V 0 8A 220 240V STEP 1 Turning counter clockwise use a screwdriver to remove the fuse case cover STEP 2 Remove the fuse and fuse case Remove the used fuse from the fuse case Insert the new fuse STEP 3 Insert the fuse case into the fuse holder Turning cloc...

Страница 25: ...of the following symbols are included in the labeling for this device Definitions accompany each symbol Please read the Operations Manual Carefully before using this device Electrical Safety Class 1 Type BF Manufacturer s ID Label 120V Manufacturer s ID Label 220V ...

Страница 26: ...ely to cause any interference in nearby electronic equipment RF emissions CISPR 11 EN55011 Class A The Dynatron 900 and accessories is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions...

Страница 27: ...al or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Compliant Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 ...

Страница 28: ...ot apply to all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environm...

Страница 29: ...rference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Dynatron 900 and accessories as recommended below according to the maximum power of the communications equipment Separation distance meters according to frequency of transmitter Rated maximum output power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 0...

Страница 30: ... of the problem with the product Write the return authorization number obtained from Dynatronics on the shipping label SHIPPING COSTS Within the first 30 days of the warranty period Dynatronics will pay all necessary shipping costs associated with obtaining service under this warranty After the first 30 days of the warranty period the owner is responsible for all costs associated with shipping the...

Страница 31: ...Dynatron 900 27 ...

Страница 32: ...n contained in the operator s manual for this device o I have received in service training from my dealer and or Dynatronics for this device IMPORTANT If there is anything about the operation or use of your Dynatron device that you do not understand contact your dealer or Dynatronics for instruction As a trained medical practitioner you are solely responsible for determining appropriate applicatio...

Отзывы: