3
3 Installation
This section describes the instructions that you should abide by when installing the
centrifuge. Before moving the centrifuge, the rotor must be removed.
WARNING:
This centrifuge may be damaged if it is connected to an improper power source.
Check if the power source meets the requirements specified on the plate at the
rear of this centrifuge.
3.1 Location
1
)
Locate this centrifuge on a firm and level table, ensure the four feet of this centrifuge
stand on the table firmly. Avoid installing on the slippery table otherwise the centrifuge will
have a bigger vibration.
2
)
Ideal environment temperature is 20
℃
±5
℃
. Temperature should not be over 30
℃
.
Avoid direct sunlight to the instrument.
3
)
Keep 300 mm distance on both sides and back of this centrifuge. This is a safe
distance of the centrifuge. After installing, this distance should be marked around the
centrifuge to ensure its cooling efficiency.
4
)
Make sure there is no water or heat around the centrifuge. Otherwise, it will result in
sample’s temperature increasing or malfunction of centrifuge.
WARNING:
No any hazardous materials are within 30cm around the centrifuge to
avoid leaking due to heavy vibration of the centrifuge.
3.2 Connection of the power cord and grounding
WARNING:
Do not touch the power cord with wet hands to avoid electrical shocks.
This centrifuge must be properly grounded.
1) This centrifuge is equipped with a 3P flat plug, 1P of 3P is for grounding. Grounding can
be done by plugging the 3P plug into the outlet.
2) The outlet must have bigger current capacity than 10A (for 220V instrument) or 25A (for
110V instrument), and its earth terminal is installed properly.