10
CAUTION:
Make sure that the cover is put on the rotor and fixed securely.
Otherwise, the rotor or its cover may be dropped off while the instrument is running.
That might damage the centrifuge or the rotor.
6) Confirm the ID code of the rotor
This centrifuge can identify rotors automatically.
Each rotor is assigned with an ID code. After the rotor is identified, the optimum
temperature can be automatically obtained. And the function of over-speed protection and
the speed/RCF display can be realized.
The list of available rotor’s ID code, see Table 6.1.
Table 6.1 List of rotors
Rotor type
ID
code
Max.
speed
(rpm)
Max.
RCF
(×g)
Tube/bottle
Imbalance tolerance(*)
Mass
Imbalance
Capacity
Imbalance(**)
FA18A
14
18,000
27,070
10 ml tube
2.0 g/ tube
5mm/ tube
FA18B
36
18,000
24,990
1.5/2 ml tube
0.2 g/tube
5mm/ tube
FA15A
19
15,000
21,500
FA15B
16
15,000
21,800
50 ml tube
5.0g/ tube
5mm/ tube
FA15C
18
15,000
23,300
FA15D
17
15,000
20,900
80 ml tube
5.0g/ tube
2mm/ tube
FA12A
03
12,000
14,100
0.2 ml microtube
0.2 g/tube
5mm/ tube
FA10B
20
10,000
11,400
50 ml tube
5.0g/ tube
5mm/ tube
SW5
50
4,800
4,170
250 ml bottle
5.0g/ tube
2mm/ tube
SW5MP
55
4,800
3,120
96 well microplate
96 Deep well plate
-
-
SW4A
60
4,000
2,900
-
-
-
FA18C
01
18,000
28,978
10 ml tube
2.0 g/ tube
5mm/ tube
FA15E
06
15,000
22,600
5ml
tube
with
V
bottom
1.0g/tube
2mm/tube
FA15F
04
15,000
22,600
5ml tube with round
bottom
1.0g/tube
2mm/tube
FA12B
02
12,000
13,840
1.5/2 ml tube
1.5g/ tube
5mm/ tube
FA14C
13
14,000
20,920
50 ml tube
5.0g/ tube
5mm/ tube