![Dynamica VELOCITY 18R Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263020.webp)
14
are sold on the market have lower allowable speeds (allowable RCF) than the rotor.
Use them at the lowest allowable speed or less.
(1) Select a rotor ID.
①
Press the
button.
The rotor symbol
blinks, the rotor ID can be selected.
②
Press
or
button to select the rotor ID.
The rotor ID code is given in Table 6.1or it can be obtained by rotor inquiring function
(refer to 7.6 section).
NOTE:
This step can be omitted because all rotors are automatically recognized.
(2) Set the speed, running time, sample temperature, acceleration and deceleration rate.
①
Press
button.
Speed symbol
blinks.
When the speed unit is RCF, it indicates you can input required RCF value. Then, press
the
button again, the speed unit becomes rpm, now user can input speed value. In
the same way, user can also change rpm into RCF.
If no button is pressed after the symbol
has been blinking for 7 seconds, the
symbol will stop blinking, close the inputting mode.
②
Press
and
buttons to adjust the parameter.
The minimum speed you can set is 300 rpm with the interval of 100rpm.
The minimum RCF you can set is 100×g with the interval of 50×g.
When keep pressing
and
buttons, you can set the values in a fast mode.
There is a cycling function on
and
buttons. When pressing the
button to the maximum value, it will return to minimum value, vice verse for pressing
button.
NOTE:
The centrifuge may shake slight when it is operated under 2,600rpm, this does
not indicate any problem.
③
Press the
button to set sample temperature.
When the temperature symbol
blinks, the temperature is ready to be set.