![Dynamica VELOCITY 18R Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263040.webp)
34
16 Circuit connecting graph
The electric connecting drawing is shown as Figure 16-1. The electric system consists
of control board, drive board, display board, sensors, motor, refrigerator, and fans etc. The
fuses are placed on the drive board with signs of F1, F2, F3 and signal board with signs of
F1, F2, F3, F4 respectively, their specifications are:
Signal board: F4: 3.15A,
Ф5×20, ordinary type, used for the DC POWER protection;
220V instrument:
Drive board: F1, F2, F3: 10A,
Ф6×30, delay type, used for the compressor, the fan and
main circuit protection;
110V instrument:
Drive board: F1, F2, F3: 25A,
Ф6×30, delay type, used for the compressor, the fan and
main circuit protection;
Figure 16-1 Circuit connecting graph
Unauthorized repairs, disassembly, and other services to the centrifuge are strictly
prohibited.