![Dynamica VELOCITY 18R Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263021.webp)
15
Press
or
button to set the temperature value, the range is from -10
℃
to
40
℃
with the interval of 1
℃
.
④
Press
button to set running time.
While the time symbol
blinks, the running time can be set.
Press
or
button to set the running time, the range is from 1minute to 99
hours and 59minutes~HOLD with the interval of 1 minute.
⑤
Press
button to set acceleration rate.
While the acceleration symbol
blinks, the acceleration rate can be set.
Press
or
button to set the acceleration rate, the range is from 1 to 9 with
the interval of 1. Curve 9 is the highest rate.
⑥
Press
button to set deceleration rate.
While the deceleration symbol
blinks, the deceleration rate can be set.
Press
and
buttons to set the deceleration rate, the range is from 1 to 9
with the interval of 1. Curve 9 is the fastest rate.
5) Start running.
(1) Press
button, to check the setting values again.
Press
button, the screen displays the parameter verification screen (see the
following interface)
Figure 7-3 Verification of the operation parameters
If something is wrong with the operation parameter, press
button, the centrifuge
will return to the preparation mode for correction.
If no button is pressed for 7 seconds, the centrifuge will return to the preparation
18000
rpm
4
02
:
30
h
:
m
Continue?
(Y/
START
, N/
STOP
)
9
7