![Dynamica VELOCITY 18R Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263010.webp)
4
3.3 Horizon adjustment
1) Turn on the power switch. After initialization, open the cover and turn off the power
switch. If the machine is not connected to the power supply, user may refer to the method
explained in section 12.1, and then open the cover.
2) Put the Level on the axis crown (see the following figure), and revolve the axis crown
for at least one circle and observe whether the air bubble is always in the central of the
level to check whether the centrifuge is horizontal.
3) If the centrifuge is not horizontal, adjust the four support feet under the bottom plate
(see the following figure) until the air bubble stay in the central of level. During adjustment,
user should rotate the axis crown continuously and check the position of air bubble. To
adjust the level, unscrew anticlockwise the lock screws to loosen them. Turn the support
feet clockwise to lower the centrifuge, vice verse. (Use the wrench in the accessory box if
needed.)
4) When the instrument is horizontal, make sure four support feet are firmly standing on
the table. Any suspended support foot is not allowed. Slightly press the top of the
Adjustable support feet
Level
Air bubble
Drive shaft (crown)
Lock nut