![Dynamica VELOCITY 18R Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263037.webp)
31
SW4A
4,000rpm
2,900×g
-
-
-
FA18C
18,000rpm
28,978×g
10×10
10ml tube
Φ16×82
FA15E
15,000rpm
22,600×g
5×12
5 ml tube with V bottom
Ф16.7×60
FA15F
15,000rpm
22,600×g
5×16
5ml tube with round bottom
Ф13.5×53.5
FA12B
12,000rpm
13,840×g
2×48
1.5ml tube
Φ10.8×40.5
FA14C
14,000rpm
20,920×g
50×4
50ml TC tube
Φ29×115
FA5A
5,600rpm
4,700×g
15×12
15ml TC tube
Φ17×119
13.2 Cleaning and sterilizing tubes and bottles
1) To choose optional conditions for cleaning and sterilizing the tubes and bottles, please
refer to the following table.
Cleaning and sterilizing conditions for tubes and bottles
O: Applicable X: Inapplicable
Condition Material
PA
PC
PP
Cle
a
n
in
g
Running water
cleaning
Acidic detergent(pH5 or lower)
X
X
X
Acidic detergent (higher than pH5 )
O
O
O
Alkaline detergent(higher than pH9 )
O
X
O
Alkaline detergent(pH9 or lower)
O
O
O
Neutral detergent(pH7)
O
O
O
Warm water (up to 70
℃
)
O
O
O
Ultrasonic
cleaning
Neutral detergent (pH7)
O
O
O
S
te
ri
liz
a
ti
o
n
Autoclaving
115
℃
(0.7kg/cm
2
) 30minutes
O
O
O
121
℃
(1.0kg/cm
2
) 20 minutes
X
O
O
126
℃
(1.4kg/cm
2
) 15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet
sterilization
200-300nm
X
X
X