iv
CAUTION:
The centrifuge must be located on a firm and level table.
Be careful not to get your finger or hands caught between the door hook and the
table when closing the door.
Do not move or relocate this centrifuge while the rotor is rotating.
If dew drops found in the rotor chamber, drain out the fluid through the drain hole to
prevent the sample to get mixed up with them and prevent them from leaking into the
drive unit.
Keep the chamber clean and remove any objects before running the instrument.
Cautions on rotors:
1
)
Always check for corrosion and damages on the rotor surface before using it. Do
not use the rotor or bucket if such abnormality is found.
2
)
Do not run this centrifuge over the allowable maximum speed of the rotor, buckets,
and adapters. If their maximum speeds vary, run it at the lowest maximum speed
among them.
3
)
Do not exceed the allowable imbalance tolerance.
4
)
Make sure the tubes and bottles within their actual capacities.
5
)
Make sure all the buckets are the same type at all times.
6
)
If the rotor is provided with a cover, make sure it is tightly rotated on the rotor
before the operation.
7
)
Use recommended rotors only. If any abnormal condition occurs during operation,
stop it immediately and contact our
service representative. Notify the service
representative the error code.
Earthquakes may cause damage to the centrifuge. Contact our service
representative if abnormality observed.