RMO60TD Error Messages
8.3 Error Message "Polarity"
The Error message "Polarity" is displayed if:
the Voltage sense cables are connected with opposite polarity either on the side of a device, or the side
of test object, and the signal is higher than 60 mV.
In some cases discharge LED can light alternately.
This message is displayed during a test. The DV-
Win software displays “polarity” and continues
measuring with erroneous test results.
Note:
If the signal is lower than 60 mV the display will show result with lower accuracy.
If some of the cables are disconnected after starting the test, error message will not show up and the
received results will be erroneous.
8.4 Error Message "Change Current"
This message is displayed if during the test the voltage at the test object is too high. In this case reduce
the test current and repeat the test.
In case when test voltage is not higher than 10 V per channel but test current is still improperly high for
that resistance error message
„Overcurrent“ is shown on the display. Red LED will light continuously.
8.5 Error Message "Overheat"
This message is displayed when RMO60TD
’s operating temperature rises too high.
In this case, wait for the test set to cool down and repeat the test.
Figure 8-4:
Error message
"Overheat"
Figure 8-2:
Error message
"Change Current"
Figure 8-3:
Error message
"Overcurrent"