Tamarisk
®
640
Software ICD
37
Response Format:
Table 61 – Automatic Calibration Activity Control ACK Response Format
Byte
Value
Description
1
0x01
Start of message.
2
0x02
ACK Response
3
0x02
Parameters length.
4, 5
0x0026
AutoCal Activity Control Command ID.
6
checksum
Frame check sequence.
3.2.7 Automatic Calibration Pending Activity Query – 0x25
Description
: This command queries the camera for pending Automatic Calibration
activities, and specifies the type of AutoCal that is pending. A suggested interval for
checking for pending calibration requests is every 5 minutes during normal operation and
every one (1) minute for the first ten (10) minutes immediately following power-up.
Command Format:
Table 62 – Automatic Calibration Pending Activity Query Command Format
Byte
Value
Description
1
0x01
Start of message.
2
0x25
AutoCal Pending Activity Query command ID.
3
0x00
Parameters length.
4
checksum
Frame check sequence.
Response Format:
Table 63 – Automatic Calibration Pending Activity Query VALUE Response Format
Byte
Value
Description
1
0x01
Start of message.
2
0x45
VALUE Response
3
0x02
Parameters length.
4, 5
Unsigned int
Type of Calibration Pending:
0: None.
1: Periodic Time-out.
2: Range Change.
6
checksum