Table 49 – Field Calibrate Type IDs
ID
Description
3
1-point calibration.
4
1-point calibration with shutter disabled.
All other values reserved.
Response Format:
Table 50 – Field Calibrate Response Format
Byte
Value
Description
1
0x01
Start of message.
2
0x02
ACK Response
3
0x02
Parameters length.
4, 5
0x0027
Field Calibrate Command ID.
6
checksum
Frame check sequence.
3.2.2 Field Calibrate – Shutter Disable Set – 0x81
Description
: Enables or disables the TIM shutter. When the shutter is disabled it is forced
closed and the shutter will ignore requests to open for field calibrations. When the shutter is
enabled, it will open and respond to requests to open or close.
Command Format:
Table 51 – Field Calibrate Shutter Disable Set Command Format
Byte
Value
Description
1
0x01
Start of message.
2
0x81
Field Calibrate Shutter Disable Set Command ID.
3
0x02
Parameters length.
4, 5
Unsigned integer
0 = open and enable the shutter
1 = close and disable the shutter.
6
checksum
Frame check sequence.
Response Format:
Table 52 – Field Calibrate Shutter Disable Set Response Format
Byte
Value
Description
1
0x01
Start of message.
2
0x02