Troubleshooting
40
Operating Ins
tructions V
amos, 1
. US ed. .
FiHal lo
!!
The inspiratory measured value of the set
anesthetic gas is lower than the min. alarm limit
Check vaporizer setting
Check that the sample line (Luer Lock) is firmly
connected
FiHal hi
!!!
The inspiratory measured value of the set
anesthetic gas is higher than the max. alarm limit
Check vaporizer setting
FiIso lo
!!
The inspiratory measured value of the set
anesthetic gas is lower than the min. alarm limit
Check vaporizer setting
Check that the sample line (Luer Lock) is firmly
seated
FiIso hi
!!!
The inspiratory measured value of the set
anesthetic gas is higher than the max. alarm limit
Check vaporizer setting
FiSev lo
!!
The inspiratory measured value of the set
anesthetic gas is lower than the min. alarm limit
Check vaporizer setting
Check that the sample line (Luer Lock) is firmly
seated
FiSev hi
!!!
The inspiratory measured value of the set
anesthetic gas is higher than the max. alarm limit
Check vaporizer setting
Gas INOP
!
Gas measurement failure
Call DrägerService
Hal INOP
!
Anesthetic gas measurement failure
Call DrägerService
Iso INOP
!
Anesthetic gas measurement failure
Call DrägerService
N
2
O INOP
!
N
2
O measurement failure
Call DrägerService
Occlusion
!
Sample gas supply line or scavenging line
blocked
Water trap: check level and replace if necessary.
Sample line: check and replace if necessary.
Check scavenging line and replace if necessary.
Power fail
!
Power supply failure
Check power supply
Pulse lo
!!!
The pulse rate is below the set minimum alarm
limit
Check condition of patient
Pulse hi
!!
The pulse rate is higher than the set maximum
alarm limit
Check condition of patient
Pulse ?
!!!
No pulse signal detected for 10 seconds
Check condition of patient
Check SpO
2
sensor
Sensor INOP
!!!
Internal device failure
Call DrägerService
Sev INOP
!
Anesthetic gas measurement failure
Call DrägerService
SpO
2
lo
!!!
Functional oxygen saturation below the set
minimum alarm limit.
Check ventilation
Check O
2
concentration in breathing system and
in fresh gas
SpO
2
INOP
!
SpO
2
measurement failure
Call DrägerService
SpO
2
Sensor ?
!
No SpO
2
sensor connected
Connect suitable SpO
2
sensor.
Fault
Cause
Remedy