Operation in Measurement Mode
26
Operating Ins
tructions V
amos, 1
. US ed.
Selecting CO2 Unit of Measurement/CO2 Scale
To set alarm limits :
●
Turn or press dial knob to select main menu.
●
Turn dial knob to select the »
Alarm
« parameter.
●
Press dial knob to open the parameter menu.
●
Turn dial knob to select the alarm limit to be edited.
●
Press dial knob to confirm selection.
The selected alarm limit will be marked with an arrow.
●
Turn dial knob to set the alarm limits.
●
Press dial knob to confirm settings.
●
Press dial knob to return to the main menu.
●
Press dial knob again to return to the measurement display.
Selecting CO
2
Unit of Measurement/CO
2
Scale
On power-up, the monitor uses the following default values:
— CO
2
unit: as set during last power-on.
— 0 to 50 mmHg (0 to 6 Vol.%, kPa) for the CO
2
scale.
The following units can be selected as units of measurement
for CO
2
:
— mmHg
— Vol%
— kPa
The following ranges can be selected for the CO
2
measurement scale:
— 0 to 50 mmHg (0 to 6 Vol.%, kPa)
— 0 to 75 mmHg (0 to 10 Vol.%, kPa)
●
Turn or press dial knob to select main menu.
●
Turn dial knob to select »
CO
2
« parameter.
●
Press dial knob to open the parameter menu.
●
Turn dial knob to set units for CO
2
, and, if necessary, also
set the scale of the CO
2
waveform.
●
Press dial knob to confirm the settings.
The selected unit of measurement and the selected scale
are marked with an arrow.
●
Press dial knob to return to the main menu.
●
Press dial knob again to return to the measurement display.
The set unit will be used until it is changed by the user again.
Alarm lo hi exit
SpO2 92
HR 80 150
SpO2
Pulse
98
110
F
i
F
et
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
0.6
0.1
30
25
0.0
FiHal 1.5
4.5
FiCO2
0.7
EtCO2
7.0
CO2
Unit
mmHg
Vol%
kPa
Range 0-6
F
i
F
et
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
0-10
SpO2
Pulse
98
110
0.6
0.1
30
25
0.0
4.5
exit