Preparation
14
Operating Ins
tructions V
amos, 1
. US ed.
Installing the Water Trap
For operation with the internal battery (available option):
As soon as Vamos is connected to AC line power, it will start
charging the internal battery. For this, the monitor does not
need to be switched on.
If the battery is charged during operation of the Vamos monitor
a battery symbol and arrow are shown in the field for the
battery status to indicate the charging.
When the battery is completely discharged, the charging time
to a full charge is typically 10 hours.
If the battery is completely charged the symbol disappears.
With a fully charged battery, Vamos can be used for 1.5 hours
without AC power.
During operation with the internal battery (without AC power)
the power supply indicator light shows yellow and the advisory
message "Power fail!" appears.
A battery symbol together with the remaining capacity of the
battery is shown in the field for the battery status.
33 % (Example)
If the internal battery is completely discharged, the screen
display will disappear and the power failure alarm will sound.
If the battery is discharged to a level below 10 % the following
caution message appears:
Batt. low !!
Installing the Water Trap
●
Remove a new "WaterLock
®
" water trap from its packaging,
●
Note installation date on the field marked for this purpose
on the water trap,
●
Grip water trap by the ridged surfaces and push it into its
holder until you feel it click into place.
WARNING !
Do not use the monitor without a water trap in place, to
avoid the risk of contaminating the monitor.
CO2
SpO2
Pulse
98
110
F
i
F
et
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
0.6
0.1
30
25
0.0
4.5
10
0
0053
7
1
63
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS