Maintenance
37
Operating Ins
tructions V
amos, 1
. US ed.
Maintenance Intervals
Maintenance
Maintenance Intervals
CAUTION !
Maintenance
The device must be inspected and serviced at regular six
month intervals. A record must be kept on this preventive
maintenance. We recommend obtaining a service contract
with DrägerService through your vendor.
For repairs and in any case of malfunction of the device,
we recommend that you contact DrägerService.
WARNING !
To avoid any risk of infection, clean and disinfect the
Vamos monitor before any maintenance according to
established hospital procedures - this applies also when
returning the Vamos monitor for repair.
WARNING !
Never operate the monitor if it has suffered physical
damage or does not seem to operate properly. In this
case always refer servicing to DrägerService or factory
trained and authorized technical service personnel
WARNING !
When servicing the monitor, always use genuine Dräger
replacement parts. Dräger cannot warrant or endorse the
safe performance of third party replacement parts for use
with the Vamos monitor
WaterLock
®
water trap
Replace if fouled or if the message
Occlusion !
is displayed (if the
sample line is known to be open
and correctly routed)
or
at the latest after 4 weeks,
see page 31.
Inspection and
maintenance
Every 6 months by factory trained
and authorized technical service
personnel
Internal battery
Have the battery replaced by
DrägerService or factory trained
and authorized technical service
personnel if the unit cuts out after
operating for only a short time
after the battery had been fully
recharged.