Operation in Measurement Mode
29
Operating Ins
tructions V
amos, 1
. US ed.
Selecting parameters to measure
Calibration
Selecting parameters to measure
Monitoring and presentation of anesthetic agents, N
2
O, CO
2
values and SpO
2
, Pulse values can be switched on and off.
The default for both settings on power-up is »
on
«.
To set:
●
Turn or press dial knob to select main menu.
●
Turn dial knob to select »
Param.
« parameter.
●
Press dial knob to open the parameter menu.
●
Turn dial knob to select setting for gas measurement
parameters.
●
Press dial knob to confirm setting.
The setting is marked with an arrow.
●
Turn dial knob to select setting for SpO
2
and Pulse
measurement.
●
Press dial knob to confirm setting.
The setting is marked with an arrow.
●
Press dial knob to return to the main menu.
●
Press dial knob again to return to the measurement
display.
If a parameter is switched off, the respective values are
replaced by "
XX
". If the gases measurement is switched off,
the field for the CO
2
-curve shows "
XX
" in its center.
Calibration
SpO
2
measurement does not need to be calibrated.
Gas measurement is calibrated every 2 hours by the monitor.
During calibration, which lasts approximately 30 seconds, the
reading »
CAL
« appears in place of the gas measurement
values.
Gas measurement calibration may also be started manually
from the main menu:
●
Turn or press dial knob to select main menu.
●
Turn dial knob to select »
Cal
« in the main menu.
●
Press dial knob to confirm.
The device calibrates automatically; the next automatic
calibration is carried out again after 2 hours.
on
Gases
off
F
i
F
et
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
on
SpO2
Pulse
98
110
Param.
0.6
0.1
30
25
0.0
4.5
off
SpO2
exit