Electrical protection class
I
Skin temperature sensor
Type BF
Heating surface of mattress
Type BF
Safety standards
IEC 601-1
IEC 601-2-35 Part 2 (for mattress heater)
IEC 601-2-21 (for radiant heater)
Electromagnetic compatibility
Tested pursuant IEC 601-1-2
Interfaces (available options)
All interface signals are electrically isolated from any components in contact with patient.
Isolation test voltage is 1.5 kV.
BabyLink
interface
RS 232
Output voltage pursuant RS-232-C
Monitor cable (part no. 83 06 488)
Wiring scheme
Modem interface
RS 232
Connection of external modem for service
purposes
Wiring scheme
Both RS 232 interfaces and central alarm input are not electrically isolated from one another.
Technical Data
63
Operating Instructions Babytherm 8004/8010, 2. ed.
m
m
GND
GND
Babytherm 8004/8010
Pin
5
2
3
RxD
TxD
TxD
RxD
Monitor
Pin
5
3
2
9 pin Sub D
connector
9 pin Sub D
connector
Connector housing
GND
RTS
CTS
Pin
RxD
TxD
8
7
3
2
5