Before Next Use
●
Reassemble components, refer to pages 18 through 22.
●
Following wipe disinfection, operate unit for a few
hours without a patient, in order to eliminate any
disinfectant residues (at mattress temperature setting
of 37
°
C).
Radiant heater in "manual" mode set to heater level 3.
Before a patient is placed in the Babytherm 8004/8010:
●
Set up accessories for therapy as needed.
●
Check operational readiness, see pages 23 to 25.
●
Verify that mattress is sufficiently preheated.
Cleaning and Disinfection Procedures
– bassinet, inside and outside
– surrounding panels, inside and outside
– bed surface
– Plexiglas
cover
– mattress;
– chassis, including attachments
●
Remove visible impurities with a disposable cloth
wetted in a detergent.
●
Wipe-disinfect surfaces.
●
Wipe and dry with a clean, damp cloth.
– Silicone U-grommets
●
Wash with a detergent and rinse with clear water.
●
Bath disinfect parts. Then rinse in clear water and dry;
or
●
sterilize at 120
°
C (248
°
F).
– Phototherapy lenses
CAUTION !
Only use detergents/disinfectants with a
ph value of 7 to 9 for phototherapy lenses.
CAUTION !
Do not launder mattresses in an automatic
washing machine. Do not autoclave mattresses.
Care
Cleaning / Disinfecting / Sterilizing
Before Next Use
51