Deviation from set temperature
In case of deviations greater than
±
1
°
C between set
and measured mattress temperature values:
1
Yellow LED starts blinking.
2
Measured value of mattress temperature starts
blinking.
3
Display shows message:
Mattress temp. deviation >1 ºC
Press OK-button to acknowledge
.
●
Audible alarm starts, red central alarm light on radiant
heater starts flashing.
Audible alarm can be silenced for 10 minutes:
4
Press
G
key,
or
5
press
»OK«
key.
1
The yellow LED is now lit continuously.
●
Red central alarm light is switched off.
When the mattress temperature reading has returned to a
value within
±
1
°
C of set temperature:
– Yellow LED, red central alarm light and audible alarm
are switched off.
After the mattress heater is initially switched on:
– Audible alarm is silenced for 1 hour (warm-up).
1
Yellow LED is lit continuously.
Mattress temperature above 40
°
C
6
Red LED
»40
°
C«
starts flashing.
7
Measured value of mattress temperature starts
blinking.
8
Display shows message:
Mattress temp. above 40 ºC
Press OK-button to acknowledge
.
●
Audible alarm starts, red central alarm light on radiant
heater starts blinking.
9
Press
G
key,
or
10
press
»OK«
key.
6
The red LED is now lit continuously.
●
Red central alarm light is switched off.
NOTE:
The alarm is canceled automatically when tempe-
rature has decreased to less than 0.5 ºC from set value.
Operation
Using Mattress Heater (Babytherm 8010)
39
Operating Instructions Babytherm 8004/8010, 2. ed.
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
1
3
5
4
2
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
6
10
8
7
9