Dräger Medical AG & Co. KGaA
Babylog 8000
Error List
Display error
number
Monitoring PCB
error number
Meaning
810
Mains switch auxiliary contact is
faulty, (battery voltage > 12 V)
811
Safety relay on the Pneumatic
Control PCB is faulty.
During the boot sequence, master
CPU PCB and Monitoring PCB do
not run synchronously.
813
Valves driver monitoring on the
Pneumatic Analog PCB faulty.
Applies to software versions 3.n
and 4.n only: Air and O
2
pressure
supplies are not connected upon
power-on.
815
+27 V voltage supply is faulty.
816
Valve drive is faulty.
817
Loudspeaker or feedback is faulty.
818
Babylog 8000/8000 SC device
temperature is too high.
819
Temperature measurement is
faulty.
820
+15 V comparator is faulty.
Table 1: Device error messages
Содержание Babylog 8000SC
Страница 9: ...7 General...
Страница 10: ...8...
Страница 13: ...11 Function Description...
Страница 14: ...12...
Страница 59: ...57 Replacing Non Repairable Items...
Страница 60: ...58...
Страница 77: ...75 Schematics and diagrams...
Страница 78: ...76...
Страница 91: ...89 Error List...
Страница 92: ...90...
Страница 99: ...97 Annex Spare parts list Test List Technical Information according to EMC standard IEC EN 60601 1 2 2001...
Страница 100: ...98...
Страница 101: ...Spare parts catalogue Babylog 8000 Because you care Emergency Care OR Anesthesia Critical Care Perinatal Care Home Care...
Страница 132: ......
Страница 150: ......