Replacing Non-Repairable Items
Babylog 8000
3
Replacing the O
2
Sensor Capsule
•
Remove the slotted-head screws
1
in the cover on the right-hand side and carefully pull the cover out.
•
Pull the spent O
2
sensor capsule out of its mount.
•
Insert the new O
2
sensor capsule into the mount (the ring-shaped conductors must show towards the
cover).
•
Push the cover in place and secure it with the screws
1
.
•
Calibrate the O
2
sensor manually.
Fig. 2:
O
2
Sensor Capsule
3.1
O
2
Sensor Calibration
Calibration is carried out automatically every 24 hours during operation.
Calibration should be carried out manually after replacement of the sensor capsule.
Calibration can be carried out manually at any time.
3.2
Calibrating the O
2
Sensor After Replacement
•
Press the key
2
in the main menu.
1
1
Содержание Babylog 8000SC
Страница 9: ...7 General...
Страница 10: ...8...
Страница 13: ...11 Function Description...
Страница 14: ...12...
Страница 59: ...57 Replacing Non Repairable Items...
Страница 60: ...58...
Страница 77: ...75 Schematics and diagrams...
Страница 78: ...76...
Страница 91: ...89 Error List...
Страница 92: ...90...
Страница 99: ...97 Annex Spare parts list Test List Technical Information according to EMC standard IEC EN 60601 1 2 2001...
Страница 100: ...98...
Страница 101: ...Spare parts catalogue Babylog 8000 Because you care Emergency Care OR Anesthesia Critical Care Perinatal Care Home Care...
Страница 132: ......
Страница 150: ......