Function description
Babylog 8000
11
Sensors
11.1
Pressure Sensors
The pressure sensors function according to the piezoresistive principle. An approximately 4 mm into
4 mm silicon chip has 4 ion implanted resistors. The silicon chip is also equipped with a pressure
diaphragm.
11.2
Y-Piece with Flow Sensor (Babylog 8000/Babylog 8000 plus)
The Y-piece with integrated flow sensor is used in pediatric respiratory gas systems. This Y-piece
connects the respiratory gas hoses of Babylog 8000/Babylog 8000 plus to the patient catheter
connection.
Fig. 28:
Flow sensor
The flow sensor works according to the hot-wire flowmeter principle. It has an additional measuring wire
for direction recognition.
Inspiration
Exspiration
Hot wire flowmeter for
direction recognition
Hot wire flowmeter
Содержание Babylog 8000SC
Страница 9: ...7 General...
Страница 10: ...8...
Страница 13: ...11 Function Description...
Страница 14: ...12...
Страница 59: ...57 Replacing Non Repairable Items...
Страница 60: ...58...
Страница 77: ...75 Schematics and diagrams...
Страница 78: ...76...
Страница 91: ...89 Error List...
Страница 92: ...90...
Страница 99: ...97 Annex Spare parts list Test List Technical Information according to EMC standard IEC EN 60601 1 2 2001...
Страница 100: ...98...
Страница 101: ...Spare parts catalogue Babylog 8000 Because you care Emergency Care OR Anesthesia Critical Care Perinatal Care Home Care...
Страница 132: ......
Страница 150: ......