Test 2 With Control Pressure
Use the same test set-up as in Test 1 and proceed as follows:
Remove the bottom plate from the Babylog 8000/Babylog 8000 plus/
Babylog 8000 SC and disconnect the hose from connection
4
on the gas connection
block. Apply a pressure of 1.5 bar ±0.1 bar into this hose (see Figure below).
Fig. 4:
Test 2
4x1.5 Si nf hose
Babylog hose disconnected from
connection
4
on the gas
connection block.
PEEP/PIP-
Valve
V
a
lve blo
ck
Insp. block
Gas connection block
4
4
View: Gas connection block tilted by 90°