Replacing Non-Repairable Items
4
Replacing the Lip Seals Every 2 Years
•
Lift the toggle
1
and take out the expiratory valve.
Fig. 4:
Removing the Expiratory Valve
•
Use, for example, a spherical-head 2.5 mm screwdriver to remove the old lip seals
2
. Insert the new
lip seals and make sure they fit properly.
Fig. 5:
Location of the Lip Seals
•
Re-mount the expiratory valve. To do so, lift up the toggle and place the expiratory valve to the front
using the guide rods.
•
Lock the expiratory valve with the toggle