Page 3 of 4
Electromagnetic Immunity
Electromagnetic Immunity
Babylog 8000 plus is intended for use in the electromagnetic environment specified below. The operator
should assure that is used in such an environment.
Immunity against
IEC 60601-1-2 test level
Compliance
level (Babylog
8000 plus)
Electromagnetic environment
electrostatic
discharge, ESD
(IEC 61000-4-2)
contact discharge:
±
6 kV
air discharge:
±
8 kV
±
2, 4, 6 kV
±
2, 4, 8 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
electrical fast
transients / bursts
(IEC 61000-4-4)
power supply lines:
±
2 kV
longer input / output lines:
±
1 kV
±
2 kV
±
1 kV
Mains power quality should be that of a
typical commercial or hospital
environment.
surges on AC
mains lines
(IEC 61000-4-5)
common mode:
±
2 kV
differential mode:
±
1 kV
±
2 kV
±
1 kV
Mains power quality should be that of a
typical commercial or hospital
environment.
power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
3 A/m
3 A/m
In close vicinity to Babylog 8000 plus,
no equipment with extraordinary power
frequency magnetic fields (power
transformers, etc.) should be operated.
voltage dips and
short interruptions
on AC mains input
lines
(IEC 61000-4-11)
dip >95%, 0.5 p
eriods
dip 60%, 5 p
eriods
dip 30%, 25 p
eriods
dip >95%, 5 seconds
>95%, 0.5 per.
60%, 5 per.
30%, 25 per.
>95%, 5 sec.
Mains power should be that of a typical
commercial or hospital environment. If
operator requires continued operation
during power mains interruptions, it is
recommended to power the Babylog
8000 plus from an uninterruptible supply
or a battery.
radiated RF
(IEC 61000-4-3)
80 MHz – 2.5 GHz: 10 V/m
10 V/m
Recommended separation distance
from portable and mobile RF
transmitters with transmission power
P
EIRP
to Babylog 8000 plus including its
lines: 1.84 m *
√
P
EIRP
X1
RF coupled into
lines
(IEC 61000-4-6)
150 kHz – 80 MHz: 10 V
within
ISM bands,
3 V outside ISM bands
X2
10 V
3 V
Recommended separation distance
from portable and mobile RF
transmitters with transmission power
P
EIRP
to Babylog 8000 plus including its
lines: 1.84 m *
√
P
EIRP
X1
Information re electromagnetic immunity (IEC 60601-1-2: 2001, tables 202, 203, 204)
X1
:
For P
EIRP
the highest possible "equivalent isotropic radiated power" of the adjacent RF transmitter has
to be inserted (value in Watt). Also in the vicinity of equipment marked with the symbol
interference may occur. Field strengths from fixed, portable or mobile RF transmitters at the
location of Babylog 8000 plus should be less than 3 V/m in the frequency range from 150 kHz to 2.5
GHz and less than 1 V/m above 2.5 GHz.
X2
:
ISM bands in this frequency range are: 6.765 MHz - 6.795 MHz, 13.553 MHz - 13.567 MHz, 26.957
MHz - 27.283 MHz, 40.66 MHz - 40.70 MHz.
Содержание Babylog 8000SC
Страница 9: ...7 General...
Страница 10: ...8...
Страница 13: ...11 Function Description...
Страница 14: ...12...
Страница 59: ...57 Replacing Non Repairable Items...
Страница 60: ...58...
Страница 77: ...75 Schematics and diagrams...
Страница 78: ...76...
Страница 91: ...89 Error List...
Страница 92: ...90...
Страница 99: ...97 Annex Spare parts list Test List Technical Information according to EMC standard IEC EN 60601 1 2 2001...
Страница 100: ...98...
Страница 101: ...Spare parts catalogue Babylog 8000 Because you care Emergency Care OR Anesthesia Critical Care Perinatal Care Home Care...
Страница 132: ......
Страница 150: ......