Replacing Non-Repairable Items
Babylog 8000
•
Remove the fastening screws
9
from the pressure reducers.
•
Pull out the pressure reducers.
Fig. 16:
Supply Block
•
Replace the pressure reducers
10
for AIR and O
2
including the O-rings
11
(part number D19080) and
12
(part number M23154).
Fig. 17:
Pressure Reducers
•
Mount the pressure reducers onto the supply block.
•
Mount the supply block.
•
Re-insert the gasket and the filters (wire fabric side shows towards inside).
•
Mount the gas supply connection plate.
9