Function description
Babylog 8000
8.2.2
Expiration
The PEEP control valve Y 4.1 vents the control pressure at the expiratory valve Y 5.1. The expiratory
valve Y 5.1 opens the ventilation system. Expiration takes place through the open expiratory valve Y 5.1.
The ejector, which is driven through the open solenoid Y 1.4, supports expiration.
Fig. 11:
Venting
8.2.3
PEEP
The PEEP/PIP valve is a pressure regulating valve which generates a control pressure of -19 mbar to
+80 mbar. Control pressures are used to trigger the inspiratory and expiratory valves in order to deliver
the peak inspiratory pressure (PIP) during inspiration and the positive end-expiratory pressure (PEEP)
during expiration to the patient-side valves.
An ejector integrated into the valve allows a negative control pressure of up to -19 mbar.
8.2.4
CPAP
The control pressure generated by the PEEP control valve Y 4.1 acts on the control side of the expiratory
valve Y 5.1 and builds up a continuous positive airway pressure in the ventilation system.
6 Pressure measurement
5 Expiratory valve
4 PEEP/PIP valve
Y5.1
Y4.1
R1.1
S6.3
S6.4
P
E
E
P
From
Y1.4
Содержание Babylog 8000SC
Страница 9: ...7 General...
Страница 10: ...8...
Страница 13: ...11 Function Description...
Страница 14: ...12...
Страница 59: ...57 Replacing Non Repairable Items...
Страница 60: ...58...
Страница 77: ...75 Schematics and diagrams...
Страница 78: ...76...
Страница 91: ...89 Error List...
Страница 92: ...90...
Страница 99: ...97 Annex Spare parts list Test List Technical Information according to EMC standard IEC EN 60601 1 2 2001...
Страница 100: ...98...
Страница 101: ...Spare parts catalogue Babylog 8000 Because you care Emergency Care OR Anesthesia Critical Care Perinatal Care Home Care...
Страница 132: ......
Страница 150: ......