Monitoring
94
Instructions for use Babylog 8000 plus SW 5.n
O
2
monitoring
O
2
monitoring is always activated on Babylog
8000. The function cannot be deactivated.
Calibrating the O
2
sensor
Performing calibration
1
Press the
Cal. Config.
key.
2
Press the
O
2
-Cal
key.
The device calibrates the O
2
sensor and displays a
message (A).
O
2
-Cal
(B) is displayed in the status
field.
The calibration is completed after approx.
5 minutes.
O
2
-Cal
(B) is no longer displayed in the
status field.
Hiding the message (A):
Press the
OK
key.
CAUTION
Faulty calibration
If the quality of the oxygen from the central gas
supply system is insufficient, the calibration may
be faulty.
Calibrate the O
2
sensor with calibration gas
(100 % O
2
).
20950
20953
A
B