Technical data
134
Instructions for use Babylog 8000 plus SW 5.n
Displayed measured values
Insp. Resistance
<11 mbar (<11 cmH
2
O) at 30 L/min
Exp. Resistance
<5 mbar (<5 cmH
2
O) at 30 L/min
Breathing gas humidifier
In accordance with EN ISO 8185
Resistance
<20 mbar/L/s (<20 cmH
2
O/L/s)
Insp. Resistance
12 mbar/L/s (
12 cmH
2
O/L/s)
Exp. Resistance
8 mbar/L/s (
8 cmH
2
O/L/s)
Additional functions
Emergency breathing valve
Opens the breathing system in the event of a fault
Safety valve
Opens at 90 to 110 mbar (90 to 110 cmH
2
O)
Subatmospheric pressure in
HFV
ventilation mode Airway pressure may be subatmospheric during
expiration, however it will not fall below –4 mbar
(–4 cmH
2
O) in the event of a fault.
Airway pressure measurement
Sensor
2 piezo-resistive pressure sensors
Range
–10 to 100 mbar (–10 to 100 cmH
2
O)
Zero point error
±1 mbar (±1 cmH
2
O)
Reinforcement error
±3 % of measured value
Measurement conditions
In accordance with ISO 80601-2-12
Peak Inspiratory Pressure
Pinsp
Positive end-expiratory pressure
PEEP
Mean pressure
Mean
Range
0 to 99 mbar (0 to 99 cmH
2
O)
Resolution
0 to 9.9 mbar (0 to 9.9 cmH
2
O): 0.1 mbar
(0.1 cmH
2
O)
10 to 99 mbar (10 to 99 cmH
2
O): 1 mbar
(1 cmH
2
O)
Airway pressure
Bar display (on device)
Range
–10 to 80 mbar (–10 to 80 cmH
2
O)
Resolution
2 mbar (2 cmH
2
O)