Technical data
142
Instructions for use Babylog 8000 plus SW 5.n
Essential performance characteristics
The essential performance consists in a controlled
and monitored patient ventilation with user-defined
settings for the monitoring functions
– minimum breathing gas flow,
– maximum airway pressure,
– minimum and maximum O
2
concentration in
the breathing gas,
or, if a set limit is exceeded, an appropriate alarm.
Integrated monitoring also generates an alarm in
the following situations:
– Failure of the external power supply
– Failure of the gas supply (O
2
and compressed
air)
An additional performance characteristic is the
prevention of backflow of gas into the central gas
supply system.