Instructions for use Babylog 8000 plus SW 5.n
137
Technical data
Displayed calculated measured values
Compliance
C
Range
0 to 10 mL/mbar (0 to 10 mL/cmH
2
O)
Resolution
0.01 mbar (0.01 cmH
2
O)
Accuracy
±20 %
Resistance
R
Range
10 to 999 mbar/L/s (10 to 999 cmH
2
O/L/s)
Resolution
1 mbar/L/s (1 cmH
2
O/L/s)
Accuracy
±20 %
Time constant
Tc
Range
10 to 999 ms
Resolution
1 ms
Accuracy
±30 %
Overinflation index
C
20
/C
Range
0 to 5
Resolution
0.01
Ratio of
f
:
VT
RVR
Range
0 to 1000 bpm/mL
Resolution
0.1 bpm/mL
Transport coefficient
DCO
2
Range
0 to 999 mL
2
/s
Resolution
1 mL
2
/s