Getting started
56
Instructions for use Babylog 8000 plus SW 5.n
A self-test is carried out. All LEDs light up. A
continuous tone and an alarm tone sequence
sound briefly.
After the self-test is complete, the device
information is displayed, see page 99.
Then the main screen is displayed.
The device is ready for operation after
approx. 15 seconds. If the automatic calibration of
the O
2
sensor is activated, O
2
measurement is
only available after 5 minutes.
Starting ventilation
1
Press the
Vent. Mode
key (A).
2
Press the
IPPV/IMV
key (B).
3
Press the
On
key.
4
Press the
key.
The symbol for alarm limits
(C) flashes.
Set the alarm limits, see page 84.
The
Calibrate flow sensor!
alarm message is
displayed.
Calibrate the flow sensor, see page 95.