Instructions for use Babylog 8000 plus SW 5.n
39
Assembly and preparation
The maximum load of the trolley must not exceed
100 kg (220 lbs).
The following maximum loads apply to the
individual sections:
Mounting the device on the trolley
Mounting the device on the trolley
1
Lock the brakes on the trolley and check the
brake function.
2
Tilt the device forward.
3
Insert the front tabs in the grooves in the
mounting plate.
4
Lower the device, insert the rear tabs in the
grooves in the mounting plate, and secure the
device with knurled screws at the rear.
22547
Range
Maximum load Example
A
Device
holder
40 kg (88 lbs)
(of which max.
5 kg (11 lbs) on
each lateral
standard rail)
Device, patient
monitor with
holder, hinged
arm
B
Holder
10 kg (22 lbs)
Breathing gas
humidifier or
medication
nebulizer
C
Base plate 50 kg (110 lbs)
Ambient air
compressor
5°
A
B
C
WARNING
Risk of personal injury and damage to
equipment
If the device is not mounted securely to the
trolley, it may fall off the trolley.
Mount the device securely. Check for secure
fit.
20901