51
Maintenance
Dry gas calibration
For use with devices with the "Standard" mouthpiece (Figure
II)
:
●
Connect the mouthpiece of the Dräger Alcotest 7510 to the outlet of the simulator using
an adapter.
For use with devices with the "Classic" mouthpiece (Figure
I)
:
●
Insert the adapter directly into the mouthpiece receptacle.
●
Connect the Dräger Alcotest 7510 to the calibration gas bottle using a hose
(maximum length 1 m).
1
Coupling, Order No.68 09 263
2
Adapter with water-trap, Order No.68 05 473
3
Hose (Viton 5 x 1.5), Order No.12 03 150
4
Alcotest control valve, Order No.68 10 693
5
Ethanol calibration gas bottle 2AL 34 Liter, Order No.68 10 920
Ethanol calibration gas bottle 6D 103 Liter, Order No.68 10 695
Items 1 to 3 are components of the calibration set, Order No.83 17 927.
0022399
7_0
1
.eps
13,3l/min
1 2
3
4
5
2
I
II
3