52
Maintenance
Implementation
Implementation of the calibration as described in the menu on page 48.
●
After successful calibration, an accuracy check must be conducted using the menu item
» Accuracy check «
●
The
» Gas type «
settings remain effective.
●
The
» Ambient pressure «
settings remain effective, unless the device is equipped with an
ambient pressure sensor.
The result of the accuracy check must be compared to the check value.