46
The main menu
– » Settings « Menu
» BAC limits «
"PASS, ALERT, FAIL" boundaries for the display modes PAF, PF and ZPWF.
Selection and entry with the
,
and
keys.
The following display shows the interrelationship between the limits and the display of the
results, using the factory setting as an example.
Default settings:
» Blow criteria «
Setting of the minimum breath volume and the minimum blow time for a valid breath sample.
After selection of the function, entry in liters (L) with the
,
and
keys.
Factory setting is 1.2 L and 2 seconds.
Limit / Example
0
< 0.10 mg/L
< 0.24 mg/L < 0.48 mg/L
≥
0.48 mg/L
Display
mode
PF
PASS
FAIL
HIGH
PAF
PASS
ALERT
FAIL
HIGH
ZPWF
ZERO
PASS
WARN
FAIL
HIGH
Alcohol concentration
➛
—
PASS:
—
ALERT:
—
FAIL:
—
HIGH:
< 0.10 mg/L
< 0.24 mg/L
< 0.48 mg/L
≥
0.48 mg/L
M
OK
M
OK