50
Maintenance
Calibration with the Dräger-Mark II Simulator
Prepare the simulator for operation.
For use with devices with the "Standard" mouthpiece (Figure
II)
:
●
Connect the mouthpiece of the Dräger Alcotest 7510 to the outlet of the simulator using
an adapter.
For use with devices with the "Classic" mouthpiece (Figure
I)
:
●
Insert the adapter directly into the mouthpiece receptacle.
—
The connection (
3
) must be as short as possible (<5 cm) in order to prevent condensation.
1
Coupling, Order No.68 09 263
2
Adapter with water-trap, Order No.68 05 473
3
Hose (Viton 5 x 1.5), Order No.12 03 150
4
Dräger Mark IIA Simulator, Order No.45 40 088
for this
Ethanol standard solution 1.0 ‰, Order No. 67 28 838
Items 1 to 3 are components of the calibration set, Order No.83 17 927.
0032399
7_0
1
.eps
1 2
2
I
II
3
3
4