Testing the non-return valve in the expiratory valve
•
Using the syringe, slowly (permissible pressure change less than 1 mbar/s) build
up a pressure of 2.5 mbar ±0.5 mbar at the outlet of the expiratory valve.
The volume injected through the syringe in one minute must not exceed 35 mL.
7.4
Testing the emergency expiratory valve
•
Operate the unit with the test lung fitted and with the following settings:
IPPV mode
Frequency: f = 6 1/min
Tidal volume: Vt = 0.5 L/min
End-expiratory pressure: PEEP = 5 mbar
Oxygen concentration: O
2
= 21 vol.%
Inspiratory time: Tinsp = 5 s
Airway pressure limit: Paw high = 10 mbar above peak pressure
•
Seal the expiratory hose.
The unit builds up a pressure up to the Paw high limit, then the pressure is reduced
momentarily to less than 6 mbar through the emergency expiratory valve.
Syringe
Pressure gauge
Expiratory valve
Содержание Savina
Страница 2: ......
Страница 6: ...All rights reserved Copyright reserved ENK5664900IECIVZ fm 04 01 06 IV Contents ...
Страница 7: ...1 General ...
Страница 8: ...2 ...
Страница 11: ...5 Function Description ...
Страница 12: ...6 ...
Страница 29: ...23 Maintenance Procedures ...
Страница 30: ...24 ...
Страница 45: ...39 Schematics and Diagrams ...
Страница 46: ...40 ...
Страница 63: ...Annex Parts catalog Test List Technical Information ...
Страница 64: ......
Страница 66: ...2 ...
Страница 91: ......