Savina
Maintenance procedures
18. Remove the valve block.
19. Remove the NIST connector mount screws (
Figure 15
).
Figure 15
Removing the NIST connector mount
20. Remove the NIST connector mount.
21. Remove the filter and sealing ring.
22. Insert the new filter and sealing ring.
23. Reassemble the unit.
Before removing the filter and the seal, make a note of their fitting posi-
tions.
Make sure not to kink any hoses during assembly.
When mounting the rear panel, make sure the hoses point to the left. Oth-
erwise the hoses could buckle (which could lead to error message "O2
measurement").