Savina
Function Description
1
General
Savina is a long-term ventilator unit designed for patients with a tidal volume
of 50 mL and above, and for the following applications:
–
Intensive care unit
–
Recovery room,
–
Secondary transportation from hospital to hospital,
–
Patient transfer within the hospital,
–
Transfer flights.
1.1
Working principle
Savina generates the compressed air for ventilation with a blower, which
means Savina is able to ventilate without any connection of compressed air
or oxygen. A controllable valve is switched in parallel with the blower to regu-
late the pre-set ventilation parameters. The valve opens or closes according
to the pre-set ventilation parameters.
To increase the oxygen concentration in the ventilatory gas, an oxygen con-
centrator, such as Permox SilentCare, can be connected. If compressed oxy-
gen is connected to Savina, the oxygen concentration of the ventilating gas
can be precisely metered. The oxygen concentration is then metered inter-
nally with sensors, the associated electronics, and a bank of valves.
1.2
Monitoring
Savina has the following monitoring system:
Measured values
Additional information
Measurement of the airway pressure
(Paw)
Pressure curve is pT1-filtered. (T1 =
100 ms)
The bargraph is additionally
smoothed.
Measurement of insp. and exp. flow
---
Measurement of peak flow (Flow-
Peak)
---
Measurement of the inspiratory O2
concentration (FiO2)
PT1-filtered T1 = 2 s
Measurement of airway peak pres-
sure (Ppeak)
---
Measurement of plateau pressure
(Pplat)
---
Measurement of airway mean pres-
sure (Pmean)
---.
Measurement of the positive end-
expiratory pressure (PEEP)
---
Measurement of the expiratory vol-
ume (Vte)
Using the phase information from
the ventilation control, the measured
value is determined over one man-
datory breath.
Measurement of the expiratory
minute volume (MV)
Determined over 1 minute and fil-
tered.
Measurement of the spontaneous
minute volume (MVspon)
Determined over 1 minute and fil-
tered.
Содержание Savina
Страница 2: ......
Страница 6: ...All rights reserved Copyright reserved ENK5664900IECIVZ fm 04 01 06 IV Contents ...
Страница 7: ...1 General ...
Страница 8: ...2 ...
Страница 11: ...5 Function Description ...
Страница 12: ...6 ...
Страница 29: ...23 Maintenance Procedures ...
Страница 30: ...24 ...
Страница 45: ...39 Schematics and Diagrams ...
Страница 46: ...40 ...
Страница 63: ...Annex Parts catalog Test List Technical Information ...
Страница 64: ......
Страница 66: ...2 ...
Страница 91: ......