![Dräger Medical Savina Скачать руководство пользователя страница 82](http://html.mh-extra.com/html/drager-medical-0/savina/savina_manual_2527053082.webp)
5664.900 Savina
Version 4.0
Test List
A
ll right
s reserved. Co
pyright
reserved.
K
5
6
649
00
TL1.
fm
16
.12
.05
4
5
Accompanying documents check
Instructions for Use manual is available
Medical product logbook is available (applies to Germany only)
6
Safety checks
•
Check the power cable for possible damage.
•
The fuse links are as specified on the rear panel.
6.1
Test of electrical safety to VDE 0751
6.1.1
Protective earth conductor resistance
•
Measure the protective earth conductor resistance according to VDE 0751.
The protective conductor resistance should be less than/equal to 0.3 ohms.
6.1.2
Equivalent unit leakage current
•
Measure the equivalent device leakage current according to VDE 0751.
The equivalent leakage current is less than/equal to 1000
P
A
6.1.3
Equivalent patient leakage current
•
Measure the equivalent patient leakage current according to VDE 0751.
The equivalent patient leakage current is less than/equal to 5000
P
A
Содержание Savina
Страница 2: ......
Страница 6: ...All rights reserved Copyright reserved ENK5664900IECIVZ fm 04 01 06 IV Contents ...
Страница 7: ...1 General ...
Страница 8: ...2 ...
Страница 11: ...5 Function Description ...
Страница 12: ...6 ...
Страница 29: ...23 Maintenance Procedures ...
Страница 30: ...24 ...
Страница 45: ...39 Schematics and Diagrams ...
Страница 46: ...40 ...
Страница 63: ...Annex Parts catalog Test List Technical Information ...
Страница 64: ......
Страница 66: ...2 ...
Страница 91: ......