Test of electrical safety to IEC 60601-1
6.2.1
Protective earth conductor resistance
•
Measure the protective earth conductor resistance according to IEC 60601-1.
The protective earth conductor resistance should be less than/equal to 0.2 ohms.
.
6.2.2
Earth leakage current
•
Measure the earth leakage current according to IEC 60601-1.
Normal condition (N.C.
)
: I less than or equal to 500
P
A
.
Single fault condition (S.F.C.): Neutral conductor interrupted: I less than or equal to
1000
P
A.
6.2.3
Patient leakage current
•
Measure the patient leakage current according to IEC 60601-1.
Normal condition (N.C.): I less than or equal to 100
P
A
.
Single fault condition (S.F.C.): Neutral conductor interrupted: I less than or equal to
500
P
A
.