Savina
Maintenance procedures
3
Replacing the dust
filter set
1. Press and hold the two tabs (
Figure 4
-1).
2. Lift and remove the filter cover (
Figure 4
-2).
Figure 4
Removing the filter cover
3. Pull the contaminated dust filters out of the filter cover (
Figure 5
-1).
Figure 5
Dust filter
4. Install new dust filters.
5. Dispose of the contaminated dust filters with the normal household
waste.
6. Insert the two filter cover noses into the rear panel (
Figure 6
-1).
7. Place the power cable under the filter cover (
Figure 6
-2).
8. Push the tabs into the mount until they engage (
Figure 6
-3).