Maintenance procedures
Figure 21
Rechargeable batteries
11. Fit new rechargeable batteries.
Figure 22
Position of hoses
12. Reassemble the unit.
13. Check the unit according to the Test List and the Instructions for Use
manual.
When mounting the rear panel, make sure the hoses point to the left. Oth-
erwise the hoses could buckle (which could lead to error message "O2
measurement").
Inadequate voltage. The new rechargeable batteries are not suffi-
ciently charged. To charge the rechargeable batteries, leave the Sav-
ina connected to the AC power supply for at least 10 hours (it is not
necessary to switch on the Savina).