Maintenance procedures
Savina
Figure 13
Removing the inspiratory flow sensor
13. Remove large, transparent hose (which connects valve block to plug-in
unit) from the plug-in unit.
14. Pull out the valve block a little and turn it 90° counter-clockwise (
Figure
14
).
Figure 14
Turning the valve block
15. Mark the hoses leading to the valve block.
16. Remove the hoses.
17. Remove the connector of the O2 Valve PCB.
Before removing the hoses from the valve block, mark them and make a
note of their fitting positions.