Preuse Checkout
4
4-5
Part Number: 4116841-001 Rev: E
Narkomed MRI-2 Operator’s Manual
Scavenger
System
10. * Verify the performance of the scavenger system.
To test the open reservoir scavenger system:
a. Make sure a 19 mm scavenger hose is connected between the
ventilator relief valve and the scavenger interface.
b. Make sure a 19 mm scavenger hose is connected between the APL
valve on the absorber and the 19 mm hose terminal on the back of
the absorber pole.
c.
Make sure a 19 mm scavenger hose is connected between the
bottom of the absorber pole and the scavenger interface.
d. Check for moisture accumulation in the breathing and scavenger
hoses. Remove any moisture found.
e. Verify the safe performance of the open reservoir scavenger system.
With the scavenger system properly installed and operating, test
for positive and negative pressure relief.
To test for negative pressure relief:
f.
Connect a vacuum hose to the DISS threaded terminal on the left-
hand side of the scavenger (or attach a wall suction hose onto the
adapter's hose-barb fitting).
g. Short-circuit the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
h. Set the absorber's manual/automatic selector valve to BAG.
i.
Turn the APL valve control knob fully counterclockwise.
j.
Verify that the suction waste gas disposal system is active.
k. Adjust the scavenger needle valve to a flowmeter indication
between the two white lines.
l.
Close all flow control valves on the anesthesia system.
m. Occlude the absorber breathing bag terminal and observe the
breathing pressure gauge on the absorber. The gauge should
indicate only a negligible negative pressure (no lower than -0.5
cmH
2
O).
To test for positive pressure relief:
n. Perform steps A through E.
o. Adjust the scavenger needle valve to a completely closed position by
turning it fully clockwise.
p. Open the oxygen flow control valve to a flow of 10 l/min and occlude
the absorber breathing bag terminal.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...